Check the trigger mode on the trigger interface whether or not in the
“normal” or “single”, and check the trigger level is whether or not over the
wave range. If yes, please put the trigger level to the middle position or set
the trigger mode to the “Auto” position. In another hand, you could choose
the “Auto” button to set up automatically.
6. After the Acquisition is set to Averages or Display Persistence time is
set too long, the waveform refreshes slowly.
It is normal in these settings
7. The signal is displayed as ladder like waveform
(1)This phenomenon is normal. The time base maybe is too slow .you
should turn the horizontal SCALE knob to increase horizontal resolution to
improve the display.
(2) Maybe the display Type is set to “Vectors”, You could set it to Dots
mode to improve the display.